Are simple IMRT beams more robust against MLC error? Exploring the impact of MLC errors on planar quality assurance and plan quality for different complexity beams

作者:Wang, Jiazhou; Jin, Xiance; Peng, Jiayuan; Xie, Jiang; Chen, Junchao; Hu, Weigang*
来源:Journal of Applied Clinical Medical Physics, 2016, 17(3): 147-157.
DOI:10.1120/jacmp.v17i3.6022

摘要

This study investigated the impact of beam complexities on planar quality assurance and plan quality robustness by introducing MLC errors in intensity-modulate radiation therapy. Forty patients' planar quality assurance QA) plans were enrolled in this study, including 20 dynamic MLC DMLC) IMRT plans and 20 static MLC SMLC) IMRT plans. The total beam numbers were 150 and 160 for DMLC and SMLC, respectively. Six different magnitudes of MLC errors were introduced to these beams. Gamma pass rates were calculated by comparing error-free fluence and error-induced fluence. The plan quality variation was acquired by comparing PTV coverage. Eight complexity scores were calculated based on the beam fluence and the MLC sequence. The complexity scores include fractal dimension, monitor unit, modulation index, fluence map complexity, weighted average of field area, weighted average of field perimeter, and small aperture ratio < 5 cm(2) and < 50 cm(2)). The Spearman's rank correlation coefficient was calculated to analyze the correlation between these scores and gamma pass rate and plan quality variation. For planar QA, the most significant complexity index was fractal dimension for DMLC p = -0.40) and weighted segment area for SMLC p = 0.27) at low magnitude MLC error. For plan quality, the most significant complexity index was weighted segment perimeter for DMLC p = 0.56) and weighted segment area for SMLC p = 0.497) at low magnitude MLC error. The sensitivity of planar QA was weakly associated with the field complexity with low magnitude MLC error, but the plan quality robustness was associated with beam complexity. Plans with simple beams were more robust to MLC error.