Method for the Determination of Total Selenium in a Wide Variety of Foods Using Inductively Coupled Plasma/Mass Spectrometry

作者:Sullivan Darryl*; Zywicki Richard; Yancey Mark
来源:Journal of AOAC International, 2013, 96(4): 786-794.
DOI:10.5740/jaoacint.12-389

摘要

An optimized method for use in a high volume laboratory was developed and validated for the determination of total selenium in a wide variety of food products including infant formula. The method utilizes a single-stage, closed-vessel microwave digestion procedure with nitric acid. Upon completion of digestion, the solution is brought to volume with water. A dilution is prepared for analysis by adding methanol (MeOH) to an aliquot and then brought to volume by nitric acid diluent resulting in a 3% MeOH concentration. The amount of selenium is determined by inductively coupled plasma (ICP)/MS with the capability of removing argon-based polyatomic interferences. Experiments were conducted to validate the method by determining precision, accuracy, linearity, specificity, ruggedness, and robustness. The calculated practical reporting level for this method was 25.0 mu g/kg. The method provided an average RSD of 2.03% during analysis of a wheat flour reference material and 2.73% during analysis of a powdered infant formula. The accuracy of the method for these materials was 98.6 and 105%, respectively. The method has also been used successfully on a variety of foods from various categories including fruits, grains, vegetables, protein, and dairy.

  • 出版日期2013-8

全文