Use of ultrasonic pachymetry for measurement of changes in corneal thickness in mouse corneal transplant rejection

作者:Flynn T H*; Ohbayashi M; Dawson M; Siddique M; Ono S J; Larkin D F P
来源:British Journal of Ophthalmology, 2010, 94(3): 368-371.
DOI:10.1136/bjo.2009.160671

摘要

Background/aims Diagnosis of rejection in the mouse model of corneal transplantation is based on subjective judgement of loss of graft transparency. The aims of this study were to (1) evaluate a pachymetry technique to measure changes in mouse corneal thickness and (2) correlate increases in transplant thickness with clinical and histological features of rejection. Methods Orthotopic corneal allografts (C57BL/6 strain donor) and syngeneic grafts were performed in A/J mice. Graft transparency was graded and corneal thickness measured by pachymetry on alternate days. Transverse sections of donor cornea excised from eyes representative of clinical opacity grades 1-4 were prepared, photographed, graft section thickness measured and stromal graft-infiltrating cells counted. Intraobserver and interobserver variations in pachymetry were statistically tested. Results Graft thickness, as measured by pachymetry, increased with each clinical opacity grade. Thickness for opacity grades 0, 1 and 2 was less than 300 mu m in all recipients. Graft thickness for grades 3 and 4 was greater than 300 mu min all cases. For measurements up to 400 mu m, there was a good correlation between thickness as measured by in vivo pachymetry and in histopathological sections. The mean interobserver bias was -11.35 mu m, while the mean intraobserver bias was +3.96 mu m. Stromal cellularity increased with increasing corneal thickness up to approximately 300 mu m. Conclusion In vivo graft pachymetry provides a new and reliable way to objectively diagnose rejection in the mouse model of corneal transplantation.

  • 出版日期2010-3