摘要

Comparative measurements of the height of the nanorelief steps of the surface of a silicon wafer by methods of three-dimensional reconstruction using a scanning electron microscope and profilometry were carried out. To realize the method of three-dimensional reconstruction, an islet gold film is formed on the surface of the test sample. The AlphaStep D-600 profiler is calibrated with a KTS-4500 QS step height gauge. The coincidence of the results of measurements with the methods of three-dimensional reconstruction and profilometry is established. The possibilities of reducing the uncertainty of the measurement results by the method of three-dimensional reconstruction were analyzed.

  • 出版日期2017-6

全文