摘要

Carpal tunnel syndrome is a frequently encountered peripheral nerve disorder caused by mechanical insult to the median nerve, which may in part be a result of impingement by the adjacent digital flexor tendons. Realistic finite element (FE) analysis to determine contact stresses between the flexor tendons and median nerve depends upon the use of physiologically accurate material properties. To assess the transverse compressive properties of the digital flexor tendons and median nerve, these tissues from ten cadaveric forearm specimens were compressed transversely while under axial load. The experimental compression data were used in conjunction with an FE-based optimization routine to determine apparent hyperelastic coefficients (mu and alpha) for a first-order Ogden material property definition. The mean coefficient pairs were mu=35.3 kPa, alpha=8.5 for the superficial tendons, mu = 39.4 kPa, alpha=9.2 for the deep tendons, mu=24.9 kPa, alpha = 10.9 for the flexor pollicis longus (FPL) tendon, and mu = 12.9 kPa, alpha=6.5 for the median nerve. These mean Ogden coefficients indicate that the FPL tendon was more compliant at low strains than either the deep or superficial flexor tendons, and that there was no significant difference between superficial and deep flexor tendon compressive behavior. The median nerve was significantly more compliant than any of the flexor tendons. The material properties determined in this study can be used to better understand the functional mechanics of the carpal tunnel soft tissues and possible mechanisms of median nerve compressive insult, which may lead to the onset of carpal tunnel syndrome.

  • 出版日期2011-3-15