A magnetically levitated centrifugal blood pump with a simple-structured disposable pump head

作者:Hijikata Wataru; Shinshi Tadahiko; Asama Junichi; Li Lichuan; Hoshi Hideo; Takatani Setsuo; Shimokohbe Akira
来源:Artificial Organs, 2008, 32(7): 531-540.
DOI:10.1111/j.1525-1594.2008.00576.x

摘要

A magnetically levitated centrifugal blood pump (MedTech Dispo) has been developed for use in a disposable extracorporeal system. The design of the pump is intended to eliminate mechanical contact with the impeller, to facilitate a simple disposable mechanism, and to reduce the blood-heating effects that are caused by motors and magnetic bearings. The bearing rotor attached to the impeller is suspended by a two degrees-of-freedom controlled radial magnetic bearing stator, which is situated outside the rotor. In the space inside the ringlike rotor, a magnetic coupling disk is placed to rotate the rotor and to ensure that the pump head is thermally isolated from the motor. In this system, the rotor can exhibit high passive stiffness due to the novel design of the closed magnetic circuits. The disposable pump head, which has a priming volume of 23 mL, consists of top and bottom housings, an impeller, and a rotor with a diameter of 50 mm. The pump can provide a head pressure of more than 300 mm Hg against a flow of 5 L/min. The normalized index of hemolysis of the MedTech Dispo is 0.0025 +/- 0.0005 g/100 L at 5 L/min against 250 mm Hg. This is one-seventh of the equivalent figure for a Bio Pump BPX-80 (Medtronic, Inc., Minneapolis, MN, USA), which has a value of 0.0170 +/- 0.0096 g/100 L. These results show that the MedTech Dispo offers high pumping performance and low blood trauma.