Preparation and certification of arsenate [As(V)] reference material, NMIJ CRM 7912-a

作者:Narukawa Tomohiro; Kuroiwa Takayoshi; Narushima Izumi; Jimbo Yasujiro; Suzuki Toshihiro; Chiba Koichi*
来源:Analytical and Bioanalytical Chemistry, 2010, 397(2): 493-499.
DOI:10.1007/s00216-010-3564-4

摘要

Arsenate [As(V)] solution reference material, National Metrology Institute of Japan (NMIJ) certified reference material (CRM) 7912-a, for speciation of arsenic species was developed and certified by NMIJ, the National Institute of Advanced Industrial Science and Technology. High-purity As(2)O(3) reagent powder was dissolved in 0.8 M HNO(3) solution and As(III) was oxidized to As(V) with HNO(3) to prepare 100 mg kg(-1) of As(V) candidate CRM solution. The solution was bottled in 400 bottles (50 mL each). The concentration of As(V) was determined by four independent analytical techniques-inductively coupled plasma mass spectrometry, inductively coupled plasma optical emission spectrometry, graphite furnace atomic absorption spectrometry, and liquid chromatography inductively coupled plasma mass spectrometry-according to As (V) calibration solutions, which were prepared from the arsenic standard of the Japan Calibration Service system and whose species was guaranteed to be As(V) by NMIJ. The uncertainties of all the measurements and preparation procedures were evaluated. The certified value of As(V) in the CRM is (99.53 +/- 1.67) mg kg(-1) (k=2).

  • 出版日期2010-5