Acoustic Reciprocity of Spatial Coherence in Ultrasound Imaging

作者:Bottenus Nick*; Uestuener Kutay F
来源:IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2015, 62(5): 852-861.
DOI:10.1109/TUFFC.2014.006928

摘要

A conventional ultrasound image is formed by transmitting a focused wave into tissue, time-shifting the back-scattered echoes received on an array transducer, and summing the resulting signals. The van Cittert-Zernike theorem predicts a particular similarity, or coherence, of these focused signals across the receiving array. Many groups have used an estimate of the coherence to augment or replace the B-mode image in an effort to suppress noise and stationary clutter echo signals, but this measurement requires access to individual receive channel data. Most clinical systems have efficient pipelines for producing focused and summed RF data without any direct way to individually address the receive channels. We describe a method for performing coherence measurements that is more accessible for a wide range of coherence-based imaging. The reciprocity of the transmit and receive apertures in the context of coherence is derived and equivalence of the coherence function is validated experimentally using a research scanner. The proposed method is implemented on a commercial ultrasound system and in vivo short-lag spatial coherence imaging is demonstrated using only summed RF data. The components beyond the acquisition hardware and beamformer necessary to produce a real-time ultrasound coherence imaging system are discussed.

  • 出版日期2015-5