A Low Power Wearable Transceiver for Human Body Communication

作者:Huang Jin*; Wang Lei; Chen Lian Kang; Zhang Yuan Ting
来源:Annual International Conference of the IEEE-Engineering-in-Medicine-and-Biology-Society, 2009-09-03 to 2009-09-06.

摘要

This paper reports a low power transceiver designed for wearable medical healthcare system. Based on a novel energy-efficient wideband wireless communication scheme that uses human body as a transmission medium, the transceiver can achieve a maximum 15 Mbps data rate with total receiver sensitivity of -30 dBm. The chip measures only 0.56 mm(2) and was fabricated in the SMIC 0.18um 1P6M RF CMOS process. The RX consumes 5mW and TX dissipates 1mW with delivering power up to 10uW, which is suitable for the body area network short range application. Real-time medical information collecting through the human body is fully simulated. Architecture of the chip together with the detail characterizes from its wireless analog front-end are presented.