Dosimetry study for a new in vivo X-ray fluorescence (XRF) bone lead measurement system

作者:Nie Hulling*; Chettle David; Luo Liqiang; O'Meara Joanne
来源:Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms , 2007, 263(1): 225-230.
DOI:10.1016/j.nimb.2007.04.090

摘要

A new Cd-109 gamma-ray induced bone lead measurement system has been developed to reduce the minimum detectable limit (MDL) of the system. The system consists of four 16 mm diameter detectors. It requires a stronger source compared to the "conventional" system. A dosimetry study has been performed to estimate the dose delivered by this system. The study was carried out by using human-equivalent phantoms. Three sets of phantoms were made to estimate the dose delivered to three age groups: 5-year old, 10-year old and adults. Three approaches have been applied to evaluate the dose: calculations, Monte Carlo (MC) simulations, and experiments. Experimental results and analytical calculations were used to validate MC simulation. The experiments were performed by placing Panasonic UD-803AS TLDs at different places in phantoms that representing different organs. Due to the difficulty of obtaining the organ dose and the whole body dose solely by experiments and traditional calculations, the equivalent dose and effective dose were calculated by MC simulations. The result showed that the doses delivered to the organs other than the targeted lower leg are negligibly small. The total effective doses to the three age groups are 8.45/9.37 mu Sv (female/male), 4.20 mu Sv, and 0.26 mu Sv for 5-year old, 10-year old and adult, respectively. An approval to conduct human measurements on this system has been received from the Research Ethics Board based on this research.

  • 出版日期2007-10
  • 单位国家地质实验测试中心