A multi-wavelength streak-optical-pyrometer for warm-dense matter experiments at NDCX-I and NDCX-II

作者:Ni P A*; Bieniosek F M; Henestroza E; Lidia S M
来源:Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment , 2014, 733: 12-17.
DOI:10.1016/j.nima.2013.05.102

摘要

We report on a multi-wavelenglh streak-optical-pyrometer (SOP) developed the for warm-dense-matter (WDM) experiments at the existing NDCX-I facility and the NDCX-II facility currently being commissioned at LBNL. The SOP served as the primary temperature diagnostic in the recent NDCX-I experiments, in which an intense K+ beam was used to heat different metal samples into WDM states. The SOP consists of a spectral grating (visible and near-infrared spectral range) and a fast, high-dynamic-range optical streak camera. The instrument is calibrated absolutely with a NIST-traceable tungsten ribbon lamp and can itself be considered as an absolutely calibrated, time-resolving spectrometer. The sample temperature is determined from fitting the recorded thermal spectrum into the Planck formula multiplied by a model of emissivity. Published by Elsevier B.V.

  • 出版日期2014-1-1