Different calibration strategies for the analysis of pure copper metal by nanosecond laser ablation inductively coupled plasma mass spectrometry

作者:Traub Heike*; Czerwensky Michael; Matschat Ralf; Kipphardt Heinrich; Panne Ulrich
来源:Journal of Analytical Atomic Spectrometry, 2010, 25(5): 690-696.
DOI:10.1039/b915564c

摘要

In this work, different calibration strategies for the determination of trace elements in pure copper metal by nanosecond laser ablation ICP-MS were investigated. In addition to certified reference materials (CRMs), pellets of doped copper powder were used for calibration. The micro homogeneity of the CRMs as well as the solution-doped pellets was sufficient to use them as calibration samples in combination with a laser spot size of 200 mm. In contrast, pellets doped with analytes in solid form showed a significant heterogeneity. For most of the investigated analytes and copper CRMs the measured mass fractions were within +/- 20% of their certified values when other copper CRMs were used as calibration samples. When solution-doped powder pellets were used as calibration samples a systematic trend towards mass fractions below the certified values was observed for nearly all elements determined in the analysed CRMs. Thermal fractionation effects during the ablation of the solution-doped pellets were suspected as the extent of the fractionation depends on the irradiance, whereas fractionation is reduced at higher irradiance.

  • 出版日期2010