摘要

We demonstrate interstitial recovery of absorption and scattering coefficients using a custom optical probe and a Monte Carlo (MC)-based recovery algorithm. The probe consists of six side-firing spectroscopy fibers contained in a 1.1-mm outer diameter cladding, with each fiber having a different axial and angular position on the probe. Broadband white light is delivered by one of the fibers and is detected steady-state by the remaining fibers. These spatially and spectrally resolved data are analyzed using a MC-based fitting algorithm in order to extract the local optical properties. The technique was verified in tissue-simulating phantoms consisting of Intralipid-20% as a scatterer and either manganese meso-tetra (4-sulfanatophenyl) porphine or intact human erythrocytes as an absorber. Absorption coefficients were recovered with a mean error of 9% and scattering coefficients were recovered with a mean error of 19%, whereas the hemoglobin oxygen saturation was recovered with a mean error of 12%. These results demonstrate the feasibility of optical property recovery for situations in which surface-contact spectroscopy is not a possibility, and where only a single probe can be inserted into the tissue.

  • 出版日期2013-10

全文