摘要

This paper describes a novel technique for fabrication of micro- and nanofluidic device that consists of a carbon nanotube (CNT) and a polydimethylsiloxane (PDMS) microchannel. Single CNT was placed at desired locations using dielectrophoresis (DEP) and PDMS microchannel was constructed on the aligned CNT via photolithography and soft lithography techniques. This technique enables a CNT to be seamlessly embedded in a PDMS microchannel. Moreover, controlling the PDMS curing condition enables the construction of the device with or without a CNT (the device without CNT has a trace nanochannel in PDMS). Preliminary flow tests such as capillary effect and pressure-driven flow were performed with the fabricated devices. In the capillary effect tests, the flow stopped at the nanochannel in both devices. In the pressure-driven flow lower flow resistance was observed in the device with a CNT.

  • 出版日期2011-5-20