Auditory brainstem responses to CE-Chirp (R), stimuli for normal ears and those with sensorineural hearing loss

作者:Cho Sung Woo; Han Kyu Hee; Jang Hyun Kyung; Chang Sun O; Jung Hyunseo; Lee Jun Ho*
来源:International Journal of Audiology, 2015, 54(10): 700-704.
DOI:10.3109/14992027.2015.1043148

摘要

Objective: Evaluation of the characteristic differences between click-and CE-Chirp-evoked auditory brainstem responses (ABRs) in normal hearing and sensorineural hearing loss. Design: A prospective study. Ears with normal hearing and with sensorineural hearing loss were evaluated. Pure-tone audiometry and click-and CE-Chirp evoked ABRs exams were conducted for all ears. Visual detection levels, wave-V amplitudes, and latencies of the ABRs were assessed. Study sample: Twenty-two ears with normal hearing and 22 ears with sloping type sensorineural hearing loss were examined. Results: In normal-hearing ears, mean amplitudes were larger for CE-chirps than for clicks at all intensities until 80 dB nHL, at which the amplitudes dropped off, presumably due to upward spread of excitation. In ears with sensorineural hearing loss, however the drop-off was less significant at 80 dB nHL. Comparisons with pure-tone audiometry findings revealed ABRs to CE-Chirps to correlate at 0.5, 1, 2, and 3 kHz, and to clicks at 1, 2, 3, and 4 kHz. Conclusions: The CE-Chirp has advantages over clicks for examining normal ears. However, under high-level stimulation, these advantages are no longer present. In ears with sensorineural hearing loss, the upward spread of excitation is less prominent. The CE-Chirps results correlate significantly to low frequency audiometric findings at 0.5 kHz, while clicks do not.

  • 出版日期2015-10-3