A comparison between laser-doppler imaging and colorimetry in the assessment of scarring: "a pilot study"

作者:Peeters Wouter*; Anthonissen Mieke; Deliaert An; Van der Hulst Rene; Van den Kerckhove Eric
来源:Skin Research and Technology, 2012, 18(2): 188-191.
DOI:10.1111/j.1600-0846.2011.00552.x

摘要

Background: This cross-sectional pilot-study investigated the reproducibility of the LDI (Moor-LDI-B2; Moor Instruments) and the chromameter (Minolta chromameter CR-300) when used in scar assessment. Methods: Twenty-seven scars in 14 subjects were included between January and June 2003. One observer performed two times both measurements with 10 min apart. The intra-observer agreement is quantified by means of the intra-class correlations (ICC) and the standard errors of measurement (SEM) for both the LDI and the chromameter. Results: Ignoring one outlier, the ICC of the LDI = 0.856 and the SEM = 34.56. The chromameter shows a better reproducibility with an ICC of 0.93 and a SEM of 0.79. Conclusion: This pilot-study with a limited number of measurements shows a moderate reproducibility of the LDI compared to the chromameter measurements, in the assessment of respectively flux and redness in scars.

  • 出版日期2012-5