A high-resolution imaging x-ray crystal spectrometer for high energy density plasmas

作者:Chen Hui*; Bitter M; Hill K W; Kerr S; Magee E; Nagel S R; Park J; Schneider M B; Stone G; Williams G J; Beiersdorfer P
来源:Review of Scientific Instruments, 2014, 85(11): 11E606.
DOI:10.1063/1.4891053

摘要

Adapting a concept developed for magnetic confinement fusion experiments, an imaging crystal spectrometer has been designed and tested for HED plasmas. The instrument uses a spherically bent quartz [211] crystal with radius of curvature of 490.8 mm. The instrument was tested at the Titan laser at Lawrence Livermore National Laboratory by irradiating titanium slabs with laser intensities of 10(19)-10(20) W/cm(2). He-like and Li-like Ti lines were recorded, from which the spectrometer performance was evaluated. This spectrometer provides very high spectral resolving power (E/dE > 7000) while acquiring a one-dimensional image of the source.

  • 出版日期2014-11