摘要

Previously, cement was formulated with degraded fatigue properties (subcement) to simulate long-term fatigue in short-term cadaver tests. The present study determined the efficacy of subcement in a 'preclinical' test of a design change with known clinical consequences: the 'polished'-to-'matt' transition of the Exeter stem (revision rates for polished stems were twice those for matt stems). Contemporary stems were bead blasted to give R(a) = 1 mu m (matt finish). Matt and polished stems were compared in cadaver pairs under stair-climbing loads (three pairs of size 1; three pairs of size 3). Stem micromotion was monitored during loading. Post-test transverse sections were examined for cement damage. Cyclic retroversion decreased for polished stems but increased for matt stems (p < 0.0001). The implant size had a substantial effect; retroversion of (larger) size-3 stems was half that of size-1 stems, and polished size-3 stems subsided 2.5 times more than the others. Cement damage measures were similar and open through-cracks occurred around both stems of two pairs. Stem retroversion within the mantle resulted in stem cement gaps of 50-150 mu m. Combining information on cyclic motion, cracks, and gaps, it was concluded that this test 'predicted' higher revision rates for matt stems (it also implied that polished size-3 stems might be superior to size-1 stems).

  • 出版日期2010