Analyzing the Influence of a New Dental Implant Design on Primary Stability

作者:da Costa Valente Mariana Lima; de Castro Denise Tornavoi; Shimano Antonio Carlos; Lepri Cesar Penazzo; dos Reis Andrea Candido*
来源:Clinical Implant Dentistry and Related Research, 2016, 18(1): 168-173.
DOI:10.1111/cid.12324

摘要

BackgroundThe macrogeometry of dental implants strongly influences the primary stability and hence the osseointegration process. PurposeCompare the performance of conventional and modified implant models in terms of primary stability. Materials and MethodsA total of 36 implants (Neodent (R)) with two different formats (n=18): Alvim CM (Conical CM, O 4.3mmx10mm in length) and Titamax Ti (Cylindrical HE, O 4.0mmx11mm in length) were inserted into artificial bone blocks. Nine implants from each set were selected to undergo external geometry changes. The primary stability was quantified by insertion torque and resonance frequency using an Osstell device and the pullout test. One-way analysis of variance and Tukey's test were used for statistical evaluation. ResultsThe comparative analysis of the implants showed a significant increase of the insertion torque for the modified Conical CM implants (p=0.000) and Cylindrical HE (p=0.043); for the resonance frequency the modified Cylindrical HE showed a lower statistical mean (p=0.002) when compared to the conventional model, and in the pullout test both modified implants showed significant reduction (p=0.000). ConclusionsWithin the limitations of this study, the proposed modification showed good stability levels and advantages when compared to the conventional implants.

  • 出版日期2016-2