An amperometric biosensor array for precise determination of homocysteine

作者:Ching Congo Tak Shing*; Chang Jen Wei; Sun Tai Ping; Shieh Hsiu Li; Tsai Chun Lang; Huang Hsin Wei; Liu Wei Hao; Liu Chia Ming; Yeh Wei Chung; Li Jia Hao
来源:Sensors and Actuators B: Chemical , 2011, 152(1): 94-98.
DOI:10.1016/j.snb.2010.09.030

摘要

Homocysteine is a risk factor for coronary heart disease (CHD). This study aimed to develop a novel, simple and reliable homocysteine biosensor array (HBA) for precise determination of homocysteine. A low-cost and tiny amperometric HBA was fabricated with a D-amino acid oxidase immobilized on screen-printed carbon electrodes, with silver-silver chloride (Ag/AgCl) as reference electrode. The effects of the number of working electrodes (1, 2, 4 and 8) of the HBA, and the life-span on the response of the HBA were investigated. The HBA on the detection of homocysteine at +300 mV (versus Ag/AgCl) showed an excellent linear response range (6.4-100 mu M) and fast response time (< 10 s). Moreover, the detection sensitivity of the HBA with 8 working electrodes could be significantly increased (about 12-fold; p < 0.001) as compared to that of the HBA with 1 working electrode because of the increase of signal to noise ratio. Finally, the HBA showed good repeatability (maximal relative standard deviation = 3.8%) and long-term stability (84% of its initial value of current response after 2 months of storage). In conclusion, a simple, low-cost, reliable and tiny HBA was developed being able to precisely determine plasma homocysteine concentrations in both the normal physiological and pathological ranges.

  • 出版日期2011-2-20