摘要
A new 2D position sensitive time of flight detector for heavy ion ERD has been developed. The detector features separate time and position measurement in a straight geometry. An electrostatic lens focuses the secondary electrons ejected from a carbon foil onto a channel plate stack maintaining the position information despite the electron momentum distribution. For position readout a 2D Backgammon anode is used. A position resolution of <0.6 mm (FWHM) and a time resolution of 96 ps (FWHM) is demoristrated.
- 出版日期2016-3-15