Simultaneous Measurement of Plasma Concentrations of Proinsulin and C-Peptide and Their Ratio with a Trefoil-Type Time-Resolved Fluorescence Immunoassay

作者:De Pauw Pieter E M; Vermeulen Ilse; Ubani Ogonnaya C; Truyen Inge; Vekens Evilien M F; van Genderen Farah T; De Grijse Joeri W; Pipeleers Daniel G; Van Schravendijk Chris; Gorus Frans K*
来源:Clinical Chemistry, 2008, 54(12): 1990-1998.
DOI:10.1373/clinchem.2008.109710

摘要

BACKGROUND: When the concentrations of 2 or more substances are measured separately, their molar ratios are subject to the additive imprecisions of the different assays. We hypothesized that the cumulative error for concentration ratios of peptides containing a common sequence might be minimized by measuring the peptides simultaneously with a "trefoil-type" immunoassay. METHODS: As a model of this approach, we developed a dual-label time-resolved fluorescence immunoassay (TRFIA) to simultaneously measure proinsulin, C-peptide, and the proinsulin-C-peptide ratio (PI/C). A monoclonal antibody captures all C-peptide-containing molecules, and 2 differently labeled antibodies distinguish between proinsulin-like molecules and true C-peptide. RESULTS: The trefoil-type TRFIA was capable of measuring plasma C-peptide and proinsulin simultaneously without mutual interference at limits of quantification of 48 and 8125 pmol/L, and 2.1 and 197 pmol/L, respectively. Within-laboratory imprecision values for the trefoil-type TRFIA ranged between 8.4% and 12% for the hormone concentrations. Unlike the hormone results obtained with separate assays, imprecision did not increase when PI/C was calculated from trefoil assay results (P < 0.05). Peptide concentrations were highly correlated with results obtained in individual comparison assays (r(2) = 0.965; P < 0.0001). The total error for PI/C obtained with the trefoil-type TRFIA remained <= 25% over a broader C-peptide range than with separate hormone assays (79-7200 pmol/L vs 590-4300 pmol/L C-peptide). Preliminary data indicate little or no interference by heterophile antibodies. CONCLUSIONS: The developed trefoil-type TRFIA is a reliable method for simultaneous measurement of proinsulin, C-peptide, and PI/C and provides proof of principle for the development of other trefoil-type multiple-label immunoassays.

  • 出版日期2008-12