Development of a time-resolved fluorescence immunoassay for herpes simplex virus type 1 and type 2 IgG antibodies

作者:Liang, Qian-Ni; Zhou, Jian-Wei; Liu, Tian-Cai; Lin, Guan-Feng; Dong, Zhi-Ning; Chen, Zhen-Hua; Chen, Juan-Juan; Wu, Ying-Song*
来源:Luminescence, 2015, 30(5): 649-654.
DOI:10.1002/bio.2800

摘要

Enzyme-linked immunosorbent assays (ELISA) specific for anti-HSV glycoprotein G (gG) are most commonly used in the clinical diagnosis of HSV infection. But most of them are qualitative and with narrow detection ranges. A novel time-resolved fluoroimmunoassay (TRFIA) methodology was developed for the quantitative determination of HSV IgG in human serum. The assay was based on an indirect immunoassay format, and performed in 96-well microtiter plates. HSV-1 and HSV-2 were used as the coating antigens. Eu3+-labeled goat anti-(human IgG) polyclonal antibodies were used as tracers. The fluorescence intensity of each well was measured and serum HSV IgG levels quantified against a calibration curve. The detection range of the novel TRFIA was between 5 and 500AU/mL. Assay sensitivity was 0.568AU/mL. The intra- and inter-assay coefficients of variation were 0.59-3.63% and 3.65-6.81%, respectively. Analytical recovery, dilution tests and serum panel tests were performed using TRFIA and the results proved satisfactory. There were no statistically significant differences in sensitivity and specificity between the TRFIA and commercial ELISAs. An effective, sensitive and accurate quantitative HSV type 1 and type 2 IgG TRFIA was successfully developed and provided diagnostic value in clinical use.