We present a detailed optimization and characterization of the analytical performance of SU-8-based emitters for electrospray ionization mass spectrometry (ESI/MS). The improved SU-8 fabrication process presented here enhances patterning accuracy and reduces the time and cost of fabrication. All emitters are freestanding and enable sample delivery by both pressure-driven and spontaneous flows. The optimized emitter design incorporates a sharp, double-cantilevered tip implemented to the outlet of an SU-8 microchannel and provides highly sensitive ESI/MS detection. Moreover, the optimized design allows the use of relatively large microchannel dimensions (up to 200 x 50 mu m(2), W x h) without sacrificing the detection sensitivity. This is advantageous with a view of preventing emitter clogging and enabling reproducible analysis. The measured limits of detection for the optimized emitter design were 1 nm for verapamil and 4 nm for Glu-fibrinopeptide B with good quantitative linearities between 1 nm and 10 mu m (R-2 = 0.9998) for verapamil and between 4 nm and 3 mu m (R-2 = 0.9992) for Glu-fibrinopeptide B. The measured tip-to-tip repeatability for signal intensity was 14% relative standard deviation (RSD) (n = 3; 5 mu m verapamil) and run-to-run repeatability 4-11% RSD (n = 4; 5 gm verapamil) for all individual emitters tested. In addition, long-term stability of < 2% RSD was maintained for timescales of 30 min even under free flow conditions. SU-8 polymer was also shown to be chemically stable against most of the tested electrospray solvents.