摘要

This paper presents an experimental protocol for absolute calibration of photo-detectors. Spectral characterization is achieved by a methodology that unlike the usual line emissions-based method, hinges on the Bremsstrahlung radiation of a Soft X-Ray (SXR) tube only. Although the proposed methodology can be applied virtually to any detector, the application presented in this paper is based on Tore Supra's SXR diagnostics, which uses Silicon Surface Barrier Diodes. The spectral response of these n-p junctions had previously been estimated on a purely empirical basis. This time, a series of second-order effects, like the spatial distribution of the source radiated power or multi-channel analyser non linearity, are taken into account to achieve accurate measurements. Consequently, a parameterised physical model is fitted to experimental results and the existence of an unexpected dead layer (at least 5 mu m thick) is evidenced. This contribution also echoes a more general on-going effort in favour of long-term quality of passive radiation measurements on Tokamaks.

  • 出版日期2013-7-14
  • 单位中国地震局

全文