Measurement of Posterior Corneal Astigmatism by the I=OLMaster 700

作者:LaHood Benjamin R*; Goggin Michael
来源:Journal of Refractive Surgery, 2018, 34(5): 331-+.
DOI:10.3928/1081597X-20180214-02

摘要

PURPOSE: To provide the first description of posterior corneal astigmatism as measured by the IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany) and assess how its characteristics compare to previous measurements from other devices.
METHODS: A total of 1,098 routine IOLMaster 700 biometric measurements were analyzed to provide magnitudes and orientation of steep and flat axes of anterior and posterior corneal astigmatism. Subgroup analysis was conducted to assess correlation of posterior corneal astigmatism characteristics to anterior corneal astigmatism and describe the distribution of posterior corneal astigmatism with age.
RESULTS: Mean posterior corneal astigmatism was 0.24 <= 0.15 diopters (D), The steep axis of posterior corneal astigmatism was vertically oriented in 73.32% of measurements, Correlation between the magnitude of anterior and posterior comeal astigmatism was greatest when the steep axis of the anterior corneal astigmatism was oriented vertically (r = 0.68, P < .0001). Vertical orientation of the steep axis of anterior corneal astigmatism became less common as age increased, whereas for posterior corneal astigmatism it remained by far the most common orientation.
CONCLUSIONS: This first description of posterior corneal astigmatism measurement by the IOLMaster 700 found the average magnitude of posterior corneal astigmatism and proportion of vertical orientation of steep axis was lower than previous estimates, The IOLMaster 700 appears capable of providing enhanced biometric measurement for individualized surgical planning.

  • 出版日期2018-5