Doppler blood pressure measurement in pigs during anaesthesia

作者:Musk Gabrielle C*; Costa Renata S; Tuke Jonathan
来源:Research in Veterinary Science, 2014, 97(1): 129-131.
DOI:10.1016/j.rvsc.2014.04.011

摘要

To determine the accuracy of Doppler blood pressure measurement in anaesthetised pigs eight large white pigs (Sus scrofa) were anaesthetised with zolazepam and tiletamine as Zoletil 100 (4 mg/kg) and xylazine (2 mg/kg) by intramuscular injection followed by isoflurane in 100% oxygen. Blood pressure measurements were recorded using a Doppler probe on the radial artery and a catheter in the femoral artery connected to a transducer. The sample mean of the Doppler measurements were compared with the sample mean of the invasive systolic arterial blood pressure while accounting for the heart rate, end-tidal CO2 and temperature. The predicted error of the Doppler was greatest when pigs were hypotensive and normothermic: 11 mmHg (CI: 6-15). Doppler measurements should be interpreted with caution in anaesthetised pigs that are hypotensive and normothermic.

  • 出版日期2014-8