摘要

Electrochemical adsorptive cathodic stripping voltammetry determination of Ge(IV) using the catechol complex on a mercury-coated (MC) electrode was prepared using a glassy carbon electrode (GCE), the peak potential of which was -0.5 V vs. Ag/AgCl on MCGCE. The various parameters of the catechol concentration, its pH, and others were optimized. The linear working ranges were obtained in the concentration of 2-700 mu g L-1 Ge(IV). The relative standard deviation at the Ge(IV) concentration of 50 mu g L-1 was 1.37% (n = 15) using the optimum condition, and the detection limit was found to be 0.6 mu g L-1 (8.26 x 10(-9) M) (S/N = 3), with an adsorption time of 180 s. The Ge(IV) response was highly linear. This developed method was applied to Ge(IV) to determine the presence of rice grains.