摘要

A new method for the chemical separation and MC-ICPMS measurements of U-Th-Pa-Ra disequilibria in volcanic samples from a single sample aliquot is presented. The accuracy and precision of our techniques is assessed by replicate analyses of the synthetic U and Th isotope standards IRMM-184 and IRMM-35, and the secular equilibrium rock standards BCR-2, W-2a and TML A measure of true sample reproducibility including the errors related to sample processing and the chemical separation is obtained from analysis of multiple digestions of the rock standards. U-234/U-238, Th-230/U-238, Pa-231/U-235 and Ra-226/Th-230 activity ratios on ten separate dissolutions of USGS rock standard BCR-2, for example, reproduce to 0.4%, 1.4%, 1.1% and 2.4% (2 SD), respectively. All elements (U-Th-Pa-Ra) are separated from a single spiked sample aliquot and the elemental concentrations are measured on the same solutions as the isotopic compositions. Analytical artefacts from determination of elemental and isotopic concentrations on different sample aliquots due to sample heterogeneity are therefore avoided. In addition, our new chemical separation protocol is time-efficient and allows for rapid processing of volcanic samples with less than 25 fg of Pa and Ra.

  • 出版日期2010-10