Assessment of an Advanced Monoenergetic Reconstruction Technique in Dual-Energy Computed Tomography of Head and Neck Cancer

作者:Albrecht Moritz H*; Scholtz Jan Erik; Kraft Johannes; Bauer Ralf W; Kaup Moritz; Dewes Patricia; Bucher Andreas M; Burck Iris; Wagenblast Jens; Lehnert Thomas; Kerl J Matthias; Vogl Thomas J; Wichmann Julian L
来源:European Radiology, 2015, 25(8): 2493-2501.
DOI:10.1007/s00330-015-3627-1

摘要

To define optimal keV settings for advanced monoenergetic (Mono+) dual-energy computed tomography (DECT) in patients with head and neck squamous cell carcinoma (SCC). DECT data of 44 patients (34 men, mean age 55.5 +/- 16.0 years) with histopathologically confirmed SCC were reconstructed as 40, 55, 70 keV Mono + and M_0.3 (30 % 80 kV) linearly blended series. Attenuation of tumour, sternocleidomastoid muscle, internal jugular vein, submandibular gland, and noise were measured. Three radiologists with > 3 years of experience subjectively assessed image quality, lesion delineation, image sharpness, and noise. The highest lesion attenuation was shown for 40 keV series (248.1 +/- 94.1 HU), followed by 55 keV (150.2 +/- 55.5 HU; P = 0.001). Contrast-to-noise ratio (CNR) at 40 keV (19.09 +/- 13.84) was significantly superior to all other reconstructions (55 keV, 10.25 +/- 9.11; 70 keV, 7.68 +/- 6.31; M_0.3, 5.49 +/- 3.28; all P < 0.005). Subjective image quality was highest for 55 keV images (4.53; kappa = 0.38, P = 0.003), followed by 40 keV (4.14; kappa = 0.43, P < 0.001) and 70 keV reconstructions (4.06; kappa = 0.32, P = 0.005), all superior (P < 0.004) to linear blending M_0.3 (3.81; kappa = 0.280, P = 0.056). Mono + DECT at low keV levels significantly improves CNR and subjective image quality in patients with head and neck SCC, as tumour CNR peaks at 40 keV, and 55 keV images are preferred by observers. aEuro cent Mono + DECT combines increased contrast with reduced image noise, unlike linearly blended images. aEuro cent Mono + DECT imaging allows for superior CNR and subjective image quality. aEuro cent Head and neck tumour contrast-to-noise ratio peaks at 40 keV. aEuro cent 55 keV images are preferred over all other series by observers.

  • 出版日期2015-8