Automated electronic monitoring of circuit pressures during continuous renal replacement therapy: a technical report

作者:Zhang, Ling; Baldwin, Ian*; Zhu, Guijun; Tanaka, Aiko; Bellomo, Rinaldo
来源:Critical Care and Resuscitation, 2015, 17(1): 51-54.

摘要

Background: Automated electronic monitoring and analysis of circuit pressures during continuous renal replacement therapy (CRRT) has the potential to predict failure and allow intervention to optimise function. @@@ Methods: Current CRRT machines can measure and store pressure readings for downloading into databases and for analysis. We developed a procedure to obtain such data at intervals of 1 minute and analyse them using the Prismaflex CRRT machine, and we present an example of such analysis. @@@ Results: We obtained data on pressures obtained at intervals of 1 minute in a patient with acute kidney injury and sepsis treated with continuous haemofiltration at 2 L\hour of ultrafiltration and a blood flow of 200 mL/minute. Data analysis identified progressive increases in transmembrane pressure (TMP) and prefilter pressure (PFP) from time 0 until 33 hours or clotting. TMP increased from 104 mmHg to 313 mmHg and PFP increased from from 131 mmHg to 185 mmHg. Effluent pressure showed a progressive increase in the negative pressure applied to achieve ultrafiltration from 0 mmHg to -168 mmHg. The inflection point for such changes was also identified. Blood pathway pressures for access and return remained unchanged throughout. @@@ Conclusions: Automated electronic monitoring of circuit pressure during CRRT is possible and provides useful information on the evolution of circuit clotting.