To analyse the feasibility and accuracy of robotic aided interventions on a phantom when using a modern C-arm-mounted cone beam computed tomography (CBCT) device in combination with needle guidance software. A small robotic device capable of holding and guiding needles was attached to the intervention table. After acquiring a 3D data set the access path was planned on the CBCT workstation and shown on the intervention monitor. Then the robot was aligned to the live fluorosopic image. A total of 40 punctures were randomly conducted on a phantom armed with several targets (diameter 2 mm) in single and double oblique trajectory (n = 20 each). Target distance, needle deviation and time for the procedures were analysed. All phantom interventions (n = 40) could be performed successfully. Mean target access path within the phantom was 8.5 cm (min 4.2 cm, max 13.5 cm). Average needle tip deviation was 1.1 mm (min 0 mm, max 4.5 mm), time duration was 3:59 min (min 2:07 min, max 10:37 min). When using the proposed robot device in a CBCT intervention suite, highly accurate needle-based interventional punctures are possible in a reasonable timely manner in single as well as in double oblique trajectories. aEuro cent Percutaneous image-guided biopsy is an important contribution of modern radiology. aEuro cent A compact robotic device has been developed which may facilitate such procedures. aEuro cent Accurate needle-based interventions are possible in a timely manner. aEuro cent Complex trajectories and even deep access paths are possible.