摘要

In order to improve the spatial resolution of a confocal system, a radial birefringent pupil filter (RBPF) is introduced into a differential confocal system. RBPF consists of two polarizers with a birefringent element between them, and its pupil function is deduced from Jones matrix. The thickness and curvature radius of RBPF are optimized independently, using the first zero coordinate ratio. The pupil function is modulated by RBPF to enhance the half-width of the response function, and lateral resolution is improved when response curve is changed with the position of RBPF as well as the polarization; then axial super-resolution of the system can be guaranteed using differential confocal detection mechanism. In comparison with conventional pupil filtering technology, RBPF features high lateral resolution and can be easily produced; moreover, it also has a simple structure. Together with its low cost, RBPF provides a new way for the improvement of super-resolution of confocal system. It is indicated from theoretical analysis and preliminary experiments that the lateral resolution can be significantly improved and the measurement error is reduced by 76 nm when measuring a standard grating of period 3 mu m; the axial resolution up to 3 nm has been achieved using the optimized pupil filter. In addition to its application for measurement of a small irregular surface in a limited space, the whole differential confocal system proposed can be fitted onto a coordinate measuring machine for non-contact measurement of dimensions and surface roughness.

全文