Absolute determination of photoluminescence quantum efficiency using an integrating sphere setup

作者:Leyre S*; Coutino Gonzalez E; Joos J J; Ryckaert J; Meuret Y; Poelman D; Smet P F; Durinck G; Hofkens J; Deconinck G; Hanselaer P
来源:Review of Scientific Instruments, 2014, 85(12): 123115.
DOI:10.1063/1.4903852

摘要

An integrating sphere-based setup to obtain a quick and reliable determination of the internal quantum efficiency of strongly scattering luminescent materials is presented. In literature, two distinct but similar measurement procedures are frequently mentioned: a %26quot;two measurement%26quot; and a %26quot;three measurement%26quot; approach. Both methods are evaluated by applying the rigorous integrating sphere theory. It was found that both measurement procedures are valid. Additionally, the two methods are compared with respect to the uncertainty budget of the obtained values of the quantum efficiency. An inter-laboratory validation using the two distinct procedures was performed. The conclusions from the theoretical study were confirmed by the experimental data.

  • 出版日期2014-12