摘要

Introduction The purpose of this study was to evaluate if ultrasound derived measures of diaphragm thickening, rather than diaphragm motion, can be used to predict extubation success or failure. @@@ Methods Sixty-three mechanically ventilated patients were prospectively recruited. Diaphragm thickness (tdi) was measured in the zone of apposition of the diaphragm to the rib cage using a 7-10 MHz ultrasound transducer. The percent change in tdi between end-expiration and end-inspiration (Delta tdi%) was calculated during either spontaneous breathing (SB) or pressure support (PS) weaning trials. A successful extubation was defined as SB for > 48 h following endotracheal tube removal. @@@ Results Of the 63 subjects studied, 27 patients were weaned with SB and 36 were weaned with PS. The combined sensitivity and specificity of Delta tdi%>= 30% for extubation success was 88% and 71%, respectively. The positive predictive value and negative predictive value were 91% and 63%, respectively. The area under the receiver operating characteristic curve was 0.79 for Delta tdi%. @@@ Conclusions Ultrasound measures of diaphragm thickening in the zone of apposition may be useful to predict extubation success or failure during SB or PS trials.