A Thermal Noise Measurement System for Noise Temperature Standards in W-Band

作者:Kang Tae Weon*; Kim Jeong Hwan; Kang No Weon; Kang Jin Seob
来源:IEEE Transactions on Instrumentation and Measurement, 2015, 64(6): 1741-1747.
DOI:10.1109/TIM.2015.2398957

摘要

A W-band waveguide noise measurement system has been developed for noise temperature (NT) standards. The system consists of two standard noise sources and a waveguide radiometer. A horn-type cryogenic noise source operating at the liquid nitrogen temperature and an ambient temperature termination have been employed as standard noise sources. To measure noise power radiated from the noise sources including a device under test, a total-power waveguide radiometer has been constructed and its performance has been evaluated. Measured NT of a commercial waveguide noise source is presented with the measurement uncertainty of 0.10-0.13 dB (k = 2) in W-band.

  • 出版日期2015-6