摘要

A liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed to identify and quantify the camptothecin quaternary derivative CPT8 for application in pharmacokinetics and tissue distribution studies. Rat plasma and tissue samples were extracted with methanol by using camptothecin as the internal standard (IS). Chromatographic separation of CPT8 and the IS was achieved using a Hypersil GOLD C-18 column, with a flow rate of 1.0 mL/min followed by quantification with tandem mass spectrometry, operating with electrospray ionization in the positive ion mode and by applying multiple reaction monitoring. The MS/MS ion transitions were monitored at m/z 484.3-361.2 for CPT8 and m/z 349.0-305.2 for the IS (CPT). A calibration curve was constructed using CPT8 concentrations ranging from 2.5 ng/mL to 2500 ng/mL (r > 0.993). The efficiency of CPT8 extraction from plasma and tissue samples ranged from 91.23% to 105.4%. Intra- and inter-day precision (relative standard deviation) values were 0.21% and 7.25%, respectively. No matrix effects were observed. The freeze-thaw stability, post-extraction stability, and stability following short- and long-term storage at low temperatures ranged from 84.12% to 108.2%. The preclinical data obtained using this method is expected to facilitate future clinical investigations of CPT8.