A novel method to recover DD fusion proton CR-39 data corrupted by fast ablator ions at OMEGA and the National Ignition Facility

作者:Sutcliffe G D*; Milanese L M; Orozco D; Lahmann B; Johnson M Gatu; Seguin F H; Sio H; Frenje J A; Li C K; Petrasso R D; Park H S; Rygg J R; Casey D T; Bionta R; Turnbull D P; Huntington C M; Ross J S; Zylstra A B; Rosenberg M J; Glebov V Yu
来源:Review of Scientific Instruments, 2016, 87(11): 11D812.
DOI:10.1063/1.4960072

摘要

CR-39 detectors are used routinely in inertial confinement fusion (ICF) experiments as a part of nuclear diagnostics. CR-39 is filtered to stop fast ablator ions which have been accelerated from an ICF implosion due to electric fields caused by laser-plasma interactions. In some experiments, the filtering is insufficient to block these ions and the fusion-product signal tracks are lost in the large background of accelerated ion tracks. A technique for recovering signal in these scenarios has been developed, tested, and implemented successfully. The technique involves removing material from the surface of the CR-39 to a depth beyond the endpoint of the ablator ion tracks. The technique preserves signal magnitude (yield) as well as structure in radiograph images. The technique is effective when signal particle range is at least 10 mu m deeper than the necessary bulk material removal. Published by AIP Publishing.

  • 出版日期2016-11