An automated analyte electroextraction and preconcentration system, which was used as the front end for a liquid chromatography-electrospray mass spectrometry instrument, is described. The extraction was based on driving the anionic analytes across a polymer inclusion membrane by application of a potential of 200 V to the cell. Five milliliters (5 mL) of sample were passed through a flow-through cell at a flow rate of 0.2 mL/min containing a membrane 20 mu m thick. This consisted of 75% cellulose triacetate as base polymer, 12.5% of tris(2-ethylhexyl)phosphate as plasticizer, and 12.5% of Aliquat 336 as cationic carrier. The target analytes were enriched in 20 mu L of a stagnant acceptor solution prior to online LC/MS analysis. The performance of the system was demonstrated for the determination of chlorinated phenoxyacetic acid herbicides in spiked river water. Enrichment factors of similar to 200 were achieved with recoveries of typically 99% and precision values of typically 5%. The limit of detection (LOD) values were found to be between 0.03 ng/mL to 0.08 ng/mL.

  • 出版日期2014-9-2