摘要

To evaluate image quality and radiation dose of a state of the art cone beam computed tomography (CBCT) system and a multislice computed tomography (MSCT) system in patients with sinonasal poliposis. In this retrospective study two radiologists evaluated 57 patients with sinonasal poliposis who underwent a CBCT or MSCT sinus examination, along with a control group of 90 patients with normal radiological findings. Tissue doses were measured using a phantom model with thermoluminescent dosimeters (TLD). Overall image quality in CBCT was scored significantly higher than in MSCT in patients with normal radiologic findings (p-value: 0.00001). In patients with sinonasal poliposis, MSCT scored significantly higher than CBCT (p-value: 0.00001). The average effective dose for MSCT was 42 % higher compared to CBCT (108 mu Sv vs 63 mu Sv). CBCT and MSCT are both suited for the evaluation of sinonasal poliposis. In patients with sinonasal poliposis, clinically important structures of the paranasal sinuses can be better delineated with MSCT, whereas in patients without sinonasal poliposis, CBCT turns out to define the important structures of the sinonasal region better. However, given the lower radiation dose, CBCT can be considered for the evaluation of the sinonasal structures in patients with sinonasal poliposis. aEuro cent CBCT and MSCT are both suited for evaluation of sinonasal poliposis. aEuro cent Effective dose for MSCT was 42 % higher compared to CBCT. aEuro cent In patients with sinonasal poliposis, clinically important anatomical structures are better delineated with MSCT. aEuro cent In patients with normal radiological findings, clinically important anatomical structures are better delineated with CBCT.

  • 出版日期2015-7