Multi-center determination of galectin-3 assay performance characteristics: Anatomy of a novel assay for use in heart failure

作者:Christenson Robert H*; Duh Show Hong; Wu Alan H B; Smith Andrew; Abel Gyorgy; deFilippi Christopher R; Wang Shunguang; Adourian Aram; Adiletto Carol; Gardiner Peter
来源:Clinical Biochemistry, 2010, 43(7-8): 683-690.
DOI:10.1016/j.clinbiochem.2010.02.001

摘要

Background: Galectin-3 is a carbohydrate binding protein that plays many important regulatory roles in inflammation, immunity and cancer. Recent studies indicate that galectin-3 is a mediator of heart failure development and progression. Development of an improved assay for galectin-3 measurement was necessary for appropriate clinical assessment. Key analytical performance characteristics, the reference distribution and association of galectin-3 levels with clinical outcome in heart failure patients are defined.
Methods: A two-site ELISA test was examined for measurement matrix, imprecision, limits of blank, detection, and quantitation, as well as linearity, high-dose hook effect, storage stability, cross-reactivity with nine similar compounds, interference with 22 common medications and icterus, lipemia and hemolysis, all in accordance with CLSI guidance. Also the effects of human anti-mouse antibodies and rheumatoid factor on recovery of galectin-3 were investigated. The reference interval was determined for 1092 ostensibly healthy individuals. The association of galectin-3 concentrations with an outcome of mortality was examined in a preliminary analysis of 129 acute decompensated heart failure patients.
Results: Galectin-3 results were equivalent when measured in serum or EDTA plasma. Imprecision studies demonstrated that the total CV was <10% at a low concentration of 6 ng/mL, 7% near the mid-level concentration of 21 ng/mL, and 15% at the high level of 70 ng/mL. The limit of blank was 0.86 ng/mL, the limit of detection was 1.13 ng/mL, and the limit of quantitation was 0.96 ng/mL. The linear measurement range was 0.96-130 ng/mL and there was no high-dose hook at levels up to 500 ng/mL. No cross-reactivity with nine compounds structurally related to galectin-3 and no interference from 22 common medications, icterus or lipemia was found. Hemolysis and presence of human anti-mouse antibodies or rheumatoid factor were found to be potential sources of interference. Samples can be stored for up to 15 days at either 22-28 degrees C or 2-8 degrees C before analysis; measurements are stable after storage at -20 degrees C or -70 degrees C for at least 6 months and through six freeze-thaw cycles. The 90th, 95th and 97.5th percentile of the normal reference interval was 17.6, 20.3 and 22.1 ng/mL, respectively. Galectin-3 in the acute decompensated heart failure patients ranged from 4.0 to 75 ng/mL; using a cutpoint of 22.1 ng/mL in an unadjusted analysis, galectin-3 values were associated with an outcome of death (p=0.035). Galectin-3 is associated with severity of heart failure as indicated by NYHA Class (p-value for trend, 0.017).
Conclusion: This ELISA for galectin-3 measurement demonstrated acceptable analytical characteristics and was associated with mortality in a preliminary unadjusted analysis.

  • 出版日期2010-5