A Proposed Electroacoustic Test Protocol for Personal FM Receivers Coupled to Cochlear Implant Sound Processors

作者:Schafer Erin C*; Musgrave Elizabeth; Momin Sadaf; Sandrock Carl; Romine Denise
来源:Journal of the American Academy of Audiology, 2013, 24(10): 941-954.
DOI:10.3766/jaaa.24.10.6

摘要

Background: Current fitting guidelines from the American Academy of Audiology (Academy) support the use of objective electroacoustic measures and behavioral testing when fitting frequency modulation (FM) systems to hearing aids. However, only behavioral testing is recommended when fitting FM systems to individuals with cochlear implants (Cis) because a protocol for conducting electroacoustic measures has yet to be developed for this population. %26lt;br%26gt;Purpose: The purpose of this study was to propose and examine the validity of a newly developed, objective, electroacoustic test protocol for fitting electrically and electromagnetically coupled FM systems to Cl sound processors. Research Design: Electroacoustic measures were conducted and replicated in the laboratory with three contemporary Cl sound processors and several FM system combinations. A repeated measures design was used with four participants to examine the validity of the proposed electroacoustic test protocol. %26lt;br%26gt;Study Sample: Three contemporary Cl sound processors were tested electroacoustically in the laboratory while coupled to combinations of five FM receivers and four FM transmitters. Two adolescents using Cochlear Nucleus 5 sound processors and two adult participants using MED-EL OPUS 2 sound processors completed behavioral and subjective measures. %26lt;br%26gt;Data Collection and Analysis: Using current hearing aid practice guidelines from the Academy, electroacoustic measurements were conducted in the laboratory with the Cls and FM systems to determine transparency, where equivalent inputs to the Cl and FM microphones result in equivalent outputs. Using a hearing aid analyzer, acoustic output from the Cl sound processor was measured via monitor earphones and specialized equipment from Cl manufacturers with 65 dB SPL speech inputs (1) to the sound processor and (2) to the FM transmitter microphones. The FM gain or volume was adjusted to attempt to achieve transparency for outputs from the two input devices. The four participants completed some or all of the following measures: speech recognition in noise without and with two FM systems in a classroom, loudness ratings without and with two FM systems measures in a quiet condition in a classroom, and questionnaires. %26lt;br%26gt;Results: Transparency was achieved for most Cl and FM combinations, but most systems required adjustments to FM gain or volume relative to the manufacturer default setting. Despite adjustments to the systems, transparency was not attainable for some FM receiver and transmitter combinations.Behavioral testing in four participants provided preliminary support to the proposed electroacoustic test protocol. %26lt;br%26gt;Conclusions: Valid and reliable electroacoustic test measures may be feasible with Cls coupled to FM systems with specialized equipment from the Cl manufacturer. Advances in equipment available for electroacoustic testing with these devices as well as additional research will lend further support to this objective approach to fitting FM systems to Cls.

  • 出版日期2013-12