MR Thermometry in the Human Prostate Gland at 3.0T for Transurethral Ultrasound Therapy

作者:Ramsay Elizabeth*; Mougenot Charles; Kohler Max; Bronskill Michael; Klotz Laurence; Haider Masoom A; Chopra Rajiv
来源:Journal of Magnetic Resonance Imaging, 2013, 38(6): 1564-1571.
DOI:10.1002/jmri.24063

摘要

PurposeTo investigate the spatial, temporal, and temperature resolution of a segmented gradient echo echo-planar imaging (EPI) technique as applied to proton resonance frequency (PRF) shift thermometry at 3 T in the human prostate gland, and to determine appropriate sequence parameters for magnetic resonance imaging (MRI)-controlled transurethral ultrasound thermal therapy. Materials and MethodsEleven healthy volunteers (age range 23-58) were scanned at 3 T with a 16-channel torso coil to study the behavior of a gradient echo EPI thermometry sequence. The temperature stability and geometric distortion were assessed for 11 different parameter sets. In a further five volunteers, the prostate T2* was measured. ResultsFor all scan parameters investigated, the temperature standard deviation within the prostate was less than 1 degrees C, while the distortion was less than 1 mm. Temperature stability was best with higher TE values (up to 25 msec), larger voxel sizes and lower EPI factors, but this had to be balanced against requirements for good spatial and temporal resolution. Prostate T2* values ranged from 30-50 msec. ConclusionA good balance between temperature stability and temporal/spatial resolution is obtained with TE = 15 msec, voxel size = 1.14 mm, and EPI factor = 9, resulting in a dynamic scan time of 7.2 seconds for the nine slices. J. Magn. Reson. Imaging 2013;38:1564-1571.

  • 出版日期2013-12