A capacitive electrode with fast recovery feature

作者:Spinelli Enrique*; Haberman Marcelo; Garcia Pablo; Guerrero Federico
来源:Physiological Measurement, 2012, 33(8): 1277-1288.
DOI:10.1088/0967-3334/33/8/1277

摘要

Capacitive electrodes (CEs) allow for acquiring biopotentials without galvanic contact, avoiding skin preparation and the use of electrolytic gel. The signal quality provided by present CEs is similar to that of standard wet electrodes, but they are more sensitive to electrostatic charge interference and motion artifacts, mainly when biopotentials are picked up through clothing and coupling capacitances are reduced to tens of picofarads. When artifacts are large enough to saturate the preamplifier, several seconds (up to tens) are needed to recover a proper baseline level, and during this period biopotential signals are irremediably lost. To reduce this problem, a CE that includes a fast-recovery (FR) circuit is proposed. It works directly on the coupling capacitor, recovering the amplifier from saturation while preserving ultra-high input impedance, as a CE requires. A prototype was built and tested acquiring ECG signals. Several experimental data are presented, which show that the proposed circuit significantly reduces record segment losses due to amplifier saturation when working in real environments.

  • 出版日期2012-8