摘要

The lateral flow assay (LFA) strip sensor possesses many advantages as a diagnostic device, including the capabilities of rapid, one-step assay performance, and high throughput production. A major limitation of the sensor, however, is its difficulty in measuring a broad concentration range of target proteins, including C-reactive protein (CRP), due to the %26quot;hook effect.%26quot; In this study, we report the use of a three-line LFA strip sensor, adding an antigen line to the conventional two-line LFA sensor, for detecting CRP within a broad concentration range in human sera. We introduced an antigen line between test and control lines in the LFA sensor. The antigen line was formed by dispensing a CRP antibody solution followed by a CRP solution in nitrocellulose membrane. All other conditions were identical to those applied to the conventional LFA strip sensor. The CRP level in test samples was generated by data processing from the intensities of three lines. The strip sensor measured a linear detection range of CRP concentration from 1 ng/mL to 500 mu g/mL within 10 min, with a calculated detection range of 0.69 ng/mL-1.02 mg/mL. Using the developed three-line LFA sensor, 50 clinical samples were measured at a detection range of 0.4-84.7 mu g/mL. This novel and easy-to-use CRP sensor can be a useful tool for rapid, sensitive, and cost-effective detection of a broad physiological concentration range of CRP capabilities that are vital for various diagnostic applications.

  • 出版日期2014-11-15