A large area full-field EDXRF imaging system based on a THCOBRA gaseous detector

作者:Silva A L M*; Carvalho M L; Janssens K; Veloso J F C A
来源:Journal of Analytical Atomic Spectrometry, 2015, 30(2): 343-352.
DOI:10.1039/c4ja00301b

摘要

By taking advantage of the attractive features in terms of gain, position and energy resolution of the 2D-THCOBRA detector, a new large area Full-Field X-ray Fluorescence Imaging (FF-XRFI) system for EDXRF imaging applications was developed. The proposed FF-XRFI system has an active area of 10 x 10 cm(2) and can be used to examine macroscopic samples with a moderately good energy resolution (< 1.6 keV FWHM at 8 keV) and a suitable spatial resolution (similar to 500 mu m FWHM). This combination of characteristics allows us to record elemental distribution maps from the surface of different sample types by combining image and energy information. Two different approaches were used for X-ray optics, one based on a single pinhole and another based on a parallel multiple-hole collimator. To illustrate the system capabilities, some sample examples were imaged and studied.

  • 出版日期2015