A new manometry device for evaluating the sphincter of Oddi using a fiber-optic pressure sensor

作者:Hanzawa Takashi*; Matsunaga Tadao; Koike Tomoyuki; Kanno Atsushi; Masamune Atsushi; Iijima Katsunori; Shimosegawa Tooru*; Haga Yoichi
来源:Minimally Invasive Therapy and Allied Technologies, 2018, 27(4): 226-232.
DOI:10.1080/13645706.2017.1412701

摘要

Introduction: Although measuring the pressure of the sphincter of Oddi and the bile duct is considered to be an important examination, called Sphincter of Oddi manometry (SOM), some complications related to the SOM device remain unsolved.
Material and methods: To decrease adverse complications, we developed a 0.46 mm manometry and we performed some in vitro studies.
Results: We successfully developed a 0.46 mm SOM. The diameter is the thinnest size used in endoscopic examinations. The results of in vitro studies show the suitability as SOM.
Conclusion: This device will decrease the risks related to SOM examination. To confirm the safety and feasibility, further studies including in vivo studies will be needed.

  • 出版日期2018

全文