摘要

Some methods for decreasing a measurement error derived from a phase-shifting error for broadband light in phase-shifting low-coherence digital holography are proposed based on theoretical analysis and numerical calculations. It is well-known that an achromatic-phase shifter based on a rotating polarizer drastically decreases the error, but it is found that a small error remains according to the imperfection of the achromatic-phase shifter. It is also found that an ideal achromatic-phase shifter perfectly eliminates the error only when the light source has a symmetrical spectrum. Furthermore, it is demonstrated that a simple linear calibration method decreases the error in a narrow range of optical path differences if a light source with an asymmetrical spectrum is used. Finally, a nonlinear calibration method that can further decrease the error in a wide range of optical path differences is discussed.

  • 出版日期2015-10