Aptamer-based array electrodes for quantitative interferon-gamma detection

作者:Chen Yu*; Pui Tze Sian; Kongsuphol Patthara; Tang Kum Cheong; Arya Sunil K
来源:Biosensors and Bioelectronics, 2014, 53: 257-262.
DOI:10.1016/j.bios.2013.09.046

摘要

Present work describes the methylene blue tagged thiolated aptamer-modified gold micro-array based biosensor for specific detection of IFN-gamma. The microchips with the microelectrode array were fabricated using standard silicon microfabrication technologies, and modified with methylene blue tagged aptamer using standard gold thiol chemistry. Electrodes were characterized and tested using Cyclic Voltammetric (CV) and Square Wave Voltammetry (SQW) measurements in a standard three-electrode format at room temperature. On an aptamer modified electrode, aptamer density was estimated to be about 4.4 x 10(12) molecules/cm(2). In IFN-gamma studies, oxidation peak currents were found to decrease and more than 50% signal suppression was achieved at 500 ng/ml. Further, the magnitude of signal suppression was found to be logarithmically proportional to the IFN-gamma in the concentration range of 1-500 ng/ml, with a detection limit of 13 ng/ml (i.e. 0.8 fmol in used sample volume of 10 mu l). Biosensor showed negligible signal changes (5%) in a very high non-specific protein background, while still able to differentiate target protein IFN-gamma at 5 ng/ml. The results indicated that our sensor binds selectively to target molecules, and the non-specific binding where adsorption of BSA protein molecules may be effectively omitted from consideration.

  • 出版日期2014-3-15