摘要

In this paper a rapid method of anodic stripping voltammetry (ASV) coupled with experimental design was developed for simultaneous determination of Zn 2+, Cd2+, Pb2+, and Cu2+in gasoline samples. A microwave assisted gasoline digestion procedure for sample preparation was also established. The determination limit was found to be 0.24 μg L-1for Zn2+;8.58 ×10-4μg L-1for Cd2+;0.13 μg L-1for Pb 2+, and 0.87 μg L-1for Cu2+. It was found that the interferences of concomitant metal ions were negligible. Gasoline samples collected in the city of Sa˜o Luis (MA), Brazil were analyzed using this developed method, and the results demonstrated that this method was reliable and accurate. It is expected that this method can be used as a routine technique for metal ion determination in gasoline.

  • 出版日期2012